Robotic Bronchoscopy with Variable-Stiffness Catheter

New method for lung cancer biopsy

By Xing-Yu Chen

I have designed a teleoperated robotic bronchoscopy system, featuring a controllable variable-stiffness catheter that enhances stability and flexibility during transbronchial biopsies. The 7 DoF robotic system allows for translation, rotation, and bending of the bronchoscope; delivery and bending of the catheter; delivery and control of biopsy tools; as well as stiffness adjustment of the catheter, which adapts to the dynamic pulmonary environment to provide stable support during tissue sampling.

Robotic Bronchoscopy System with Variable-Stiffness Catheter for Pulmonary Lesion Biopsy

Crepe

System Overview

The system comprises a teleoperated surgical robot specifically designed for trans-respiratory diagnosis, incorporating a three-stage design scheme that includes a bronchoscope, novel variable-stiffness catheters, and biopsy forceps. The robot is mounted on a 5 DoF passive robotic arm to achieve the appropriate angle of intervention. Surgeons use tablets as the remote control system to teleoperate the three endoscopic instruments, inserting them through the patient’s trachea and utilizing them for precise movements. Employing a multi-operator strategy, the robot is controlled through a scheduling arrangement and weight distribution, allowing mentor surgeons and trainee surgeons to observe the same surgical site and collaboratively control the surgical instruments simultaneously. Compared to the surgeon console of the Da Vinci surgical robot, Force Dimension’s haptic devices, and other custom-made master control units, this HMI allows for quick integration of new features, making it convenient for prototyping. However, it lacks physical feedback and cannot provide similar force feedback information, which may increase the risk of operational hazards. Prolonged use of the touchscreen may also cause fatigue in the fingers or hands of the surgeon. Additionally, there are security concerns in wireless environments. Future improvement will transfer the control algorithm to a more ergonomic surgeon console.

Crepe

Endoscopic Intruments Parameters Size
Bronchoscope Outside Diameter 5.2 mm
Bronchoscope Inside Diameter 2.6 mm
Variable-Stiffness Catheter Outside Diameter 2.4 mm
Variable-Stiffness Catheter Inside Diameter 1.2 mm
Biopsy Forceps Diameter 1.0 mm

Multimodal Navigation System

Bronchoscopic images provide the most intuitive visual information, reflecting the internal condition of the lungs in real-time. However, due to their first-person perspective, these images do not reveal the bronchoscope’s position within the lungs. To address this limitation, the robot employs a multimodal navigation system, which includes bronchoscopic videos, electromagnetic (EM) sensors, and C-Arm fluoroscopy.

A 6-DoF EM sensor with position accuracy of 0.48 mm and orientation accuracy of 0.30° from Northern Digital Inc. is attached to the tip of bronchoscope to enable real-time navigation and tracking of the surgical instruments. This sensor provides information about the spatial position and orientation (roll, pitch, and yaw) of the bronchoscope. A field generator placed beside or under the patient generates an EM field of known geometry. As the EM sensor moves within this field, its spatial position and orientation are measured. The open-source software 3D Slicer is used to visualize the spatial position and orientation of the bronchoscope within lungs, as shown in Fig. 3. Hounsfield Unit (HU) values for many metals and alloys are higher than those of the human body [26] [27], making fluoroscopic X-ray particularly suitable for positioning the biopsy forceps and variable-stiffness catheter with low melting point alloy, especially during the clamping action in biopsy procedures. In the C-Arm medical imaging system, the Xray generator and detector are positioned on a C-shaped arm, directly opposite and centrally aligned with each other. The C-Arm fluoroscopy system is capable of controlled rotation, allowing for imaging from various angles and positions.

Crepe

Three-stage robotic bronchoscopy procedure

A novel three-stage robotic bronchoscopy procedure proposed in Fig. 4 involves the control of the bronchoscope, variable-stiffness catheter, and biopsy instruments. The surgery includes three types of operations: individual device control, dynamic adjustment, and biopsy procedure. In the first step, surgeons use tablets to navigate the bronchoscope robot to the target position, involving motions such as delivery, rotation, and tip bending. In the second step, the variable-stiffness catheter is heated and maintained at a safe temperature of 53°C to decrease its stiffness. Subsequently, the softened catheter is inserted through the bronchoscope’s working channel. Its dynamic stiffness can be adjusted based on factors such as heartbeat and respiration. Compared to the bronchoscope, the variable-stiffness catheter can be inserted into narrower bronchi and offers more flexible control with the driving tendon. Once the softened catheter reaches the precise location of the targeted lesion, the current to the heating resistance wire is disconnected. This discontinuation of heat allows the low melting point alloy to cool naturally, restoring the catheter’s stiffness.

Crepe